| Specifications GENERAL
 Operating Temperature Range(Datalogger):
 0 to 60°C(32 to 140°F) using alkaline battery;
 -40 to 65°C (-40 to 150°F) usinglithium battery
 Humidity: 0 to 95% RH,non-condensing
 Storage Temperature: -40 to 77°C(-40 to 170°F)
 Time Accuracy: ±100 ppm @ 75°F
 Power: 9 V alkaline or lithiumbattery
 Battery Life (average use): 20months using alkaline battery,
 40months using lithium battery
 Dimensions: 7.6 x 6.4 x 2.8 cm(3 x 2.5 x 1.1")
 Weight: 71 g (2.5 oz)
 Software: XP/Vista/7/8 (32-bit and 64-bit)
 Measurement Interval: 1 secto 18 hrs
 (selectable in 1 secondincrements)
 Computer Interface: RS-232standard;
 for computers with onlya USB port,
 use model no.OM-PL-USBS (RS-232/USBadaptor)
 
 OM-PLT2 TEMPERATUREDATALOGGER
 Channels: one internal temperaturechannel and one
 externaltemperature channel
 Sample Point Capacity:
 21,500points for single channel;
 10,750points per channel for dual channel
 Alarms: high and low alarms
 forinternal and external channel
 Temperature Range: internalsensor;
 0 to 60°C (32 to 140°F) withalkaline battery,
 -40 to 65°C (-40 to150°F) with lithium battery:
 externalsensor; -40 to 65°C (-40 to 150°F)
 Accuracy: ±0.5°C (±1.0°F)
 Resolution: 0.05°C (0.1°F)
 External Probe TemperatureRange:
 -40 to 65°C (-40 to 150°F)
 Calibration: single point offsetcalibration available throughsoftware
 
 OM-PLTT DUAL TEMPERATUREDATALOGGER
 Channels: two external
 temperature channels
 Sample Point Capacity: 21,500
 temperature points for single
 channel (1 probe); 10,750
 temperature points per channel
 for dual channel (2 probes)
 Alarms: high and low alarms
 foreach probe
 External Probe TemperatureRange:
 -46 to 150°C (-50 to 300°F)
 Accuracy: ±1°C (±2°F)
 Resolution: 0.05 °C (0.1°F)
 Probe: 76 mm (3") stainless steeltip,
 1.8 mm (6') cable
 Calibration: single point offsetcalibration available throughsoftware
 Battery Life (average use): 20months using alkaline battery,
 40months using lithium battery
 
 OM-PLTH TEMPERATURE/HUMIDITY DATALOGGER
 Channels: two internal channels(temperature and humidity);
 two calculated channels (dew pointand wet bulb)
 Sample Point Capacity: 21,500 fortemperature only;
 10,750 points fortemperature, humidity and dewpoint
 Alarms: high and low alarms
 fortemperature & humidity.
 Temperature Range: 0 to 60°C(32 to 140°F) with alkaline battery
 -40 to 65°C (-40 to 150°F) withlithium battery
 Temperature Accuracy:±1°C (±2°F)
 Resolution: 0.05°C (0.1°F)
 Relative Humidity Range:
 0 to 99.9% RH, non-condensing
 Relative Humidity Accuracy:
 ±2% full scale, from 0% to 95% RH
 for temperatures between 0 to 60°C(32 to 140°F)
 Relative Humidity Resolution:0.1% RH
 Calibration: single point offsetcalibration
 available throughsoftware; humidity is factorycalibrated
 
 OM-PLPT PRESSURE
 ANDTEMPERATURE DATALOGGER
 Channels: two external channels;
 temperature and pressure
 Sample Point Capacity:
 21,500points for temperature or pressure;
 10,750 points for both pressure andtemperature
 Alarms: high and low alarms
 fortemperature and pressure
 Temperature Range: -40 to 65°C(-40 to 150°F)
 Temperature Accuracy:±0.5°C (±1°F)
 Resolution: 0.05°C (0.1°F)
 Temperature Probe: -40 to 65°C(-40 to 150°F)
 Pressure Range: 0 to 500 psi
 Pressure Accuracy: ±3 psi
 Pressure Resolution: 0.15 psi
 Pressure Sensor: stainless steelsensor 1.8 m (6') cable,
 1/4 " NPTmale pressure port)
 Operating Temperature Range(Pressure Sensor):
 0 to 54°C(32 to 130°F)
 Calibration: single point offsetcalibration available throughsoftware;
 pressure is factorycalibrated
 
 
 |  | 
 OM-PLTC THERMOCOUPLEDATALOGGER
 Channels: two channels,one for internal ambienttemperature
 and one forexternal thermocouple input
 Measurement Capacity:10,750 points for both internal
 temperature and externalthermocouple input
 Thermocouple Types(External Channel):
 J, K, T, E, R, S (softwareselectable)
 Thermocouple TemperatureRanges:
 Type J; -79 to 732°C(-110 to 1350°F):
 Type K; -140 to 1038°C(-220 to 1900°F):
 Type T; -148 to 349°C(-235 to 660°F):
 Type E; -59 to 560°C(-75 to 1040°F):
 Type R; 0 to 1449°C(32 to 2640°F):
 Type S; 0 to 1449°C(32 to 2640°F)
 Thermocouple Connection:
 subminiature femalethermocouple jack
 Accuracy (ThermocoupleChannel):
 Type J,K,T; ±0.8°C (±1.5°F):
 Type E; ±0.6°C (±1.0°F):
 Type R,S; ±2.8°C (±5°F)
 Resolution (ThermocoupleChannel): 0.05°C (0.1°F)
 Accuracy (Internal Channel):±1°C (± 2°F)
 Resolution (Internal Channel):0.05°C (0.1°F)
 Alarms: high and low alarms
 forthermocouple and ambienttemperature
 Calibration: single point offsetcalibration available
 throughsoftware for both internal
 ambientchannel and external thermocouplechannel
 
 OM-PL420 4 TO 20 MADATALOGGER
 Channels: one 4 to 20 mA input
 Measurement Capacity:21,500 points
 Alarms: high and low alarms
 Input Current Range: 0 to 40 mA
 Current Accuracy: ±0.05 mA
 Current Resolution: 0.01 mA
 Calibration: single point offsetcalibration available throughsoftware
 
 OM-PLCV CURRENTAND VOLTAGE LOGGER
 Channels: two channels,one for AC current (measured
 using supplied current probe),and one for AC voltage
 Sample Point Capacity: 21,500points for current or voltage;
 10,750points for both current and voltageAlarms:
 high and low alarms forcurrent and voltage
 Current Range: 0 to 300 Aac
 Current Accuracy: ±5%
 Current Resolution: 0.1 Aac
 Voltage Range: 0 to 500 Vac
 Voltage Accuracy: ±1 Vac
 Voltage Resolution: 0.1 Vac
 Calibration: single point offsetcalibration available throughsoftware; current and voltage arefactory calibrated
 
 OM-PL SERIES LOGGERINTERFACE SOFTWARE
 SPECIFICATIONS
 Sampling Interval: User selectablefrom 1 second to 18 hours,
 in 1second increments
 Interface: COM (RS232) serial port.Software supports COM1 throughCOM8. Optional USB interface(model no. OM-PL-USBS).
 Alarms: Visual alarms with userselectable low
 and high thresholds forevery channel
 Alarm Memory: Up to 50 alarmevents.
 Alarm start time and alarmduration is stored
 Data Security: Locking sliding doorprevents physical access to thelogger data. Unique factory set loggerID.Data collection session counter
 Compatibility:XP/Vista/7/8 (32-bit and 64-bit)
 Minimum System Requirements:8MB of RAM, 2M
 of disk space, 800 x 600resolution, 1 free COM port or a freeUSB port
 (optional USB interface isrequired: model no.OM-PL-USBS)
 Recording Start: On door close,
 immediate after setup, or at a userspecified date and time
 Recording Mode: Stop on memoryfull,
 or continuous recording withmemory rollover
 Real Time Status: Visual indicationthrough the status LED. Userselectable
 status display frequencyand LED brightness, Software realtime status
 User Calibration: Passwordprotected calibration for each channel
 User Message:Up to 30 characters
 Logger Channels: User selectable.Disabling channels increases
 thestorage capacityof the other channels
 Units: US and Metric(user selectable units)
 Data Format: Custom formatexportable to text or Excelspreadsheet.
 Real Time Monitoring: Logger canbe permanently connected to thecomputer
 and used as a real time,multichannel sensor with loggingcapability.
 Battery does not drainwhen logger is connected to thecomputer
 |